Device product
BenchMark ULTRA and DISCOVERY ULTRA Instruments
Z-0894-2022
Product summary
- Event
- Event 89727
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 9214
- Official record key
device-enforcement:Z-0894-2022
Official wording
Reason: Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.
Code information: Serial Numbers Less-Than-Or-Equal-To 322030
Distribution pattern: US: CA, IL, PA, FL, AZ, NY, OH, CO, AR, NC, MD, AL, AK, LA, IA, MN, MA, ND, IN, TX, TN, CT, SC, GA, UT, WI, VA, SD, KY, MS, MO, DE, MI, MT, NJ, WV, OK, WA, OR, PR, DC, NM, ID, NH, NE, ME, KS, HI, NV, RI, WY, VT OUS: Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Japan, Netherlands, Norway, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, United Kingdom, Algeria, Argentina, Aruba, Bosnia-Herzegovina., Brazil, Bulgaria, Chile, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Greece, Indonesia, Ireland, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Martinique, Mexico, Monaco, Morocco, Namibia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Reunion, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Vietnam, White Russia, Zimbabwe
Derived failure modes
-
Unknown
Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.