Skip to content
Recall Observatory FDA recall evidence

Device product

Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.

Z-2693-2017

June 02, 2017

Class II

Product summary

Firm
St. Jude Medical, Inc.
Event
Event 77509
Status
Terminated
Classification
Class II
Quantity
443 units
Official record key
device-enforcement:Z-2693-2017

Official wording

Reason: The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.

Code information: AAD255.1 AAN997.1 AAP279.1 AAP328.1 AAP333.1 AAT509.1 AAT579.1 AAT582.1 AAT652.1 AAT783.1 AAT792.1 AAT795.1 AAT797.1 AAT804.1 AAT845.1 AAT902.1 AAT976.1 AAT987.1 AAT989.1 AAU003.1 AAU009.1 AAU022.1 AAU046.1 AAU078.1 AAU094.1 ABA946.1 ABM329.1 ABM630.1 ABT798.1 ABT841.1 ACE689.1 ACE771.1 ACE786.1 ACE838.1 ACE875.1 ACE900.1 ACE936.1 ACE939.1 ACE994.1 ACF065.1 ACK145.1 ACK161.1 ACK241.1 ACK424.1 ACK511.1 ACK514.1 ACK549.1 ACK559.1 ACK622.1 ACK629.1 ACK637.1 ACK699.1 ACK702.1 ACK739.1 ACK752.1 ACK803.1 ACK805.1 ACK806.1 ACK828.1 ACK855.1 ACP680.1 ACP686.1 ACP687.1 ACP700.1 ACP770.1 ACP780.1 ACP824.1 ACP833.1 ACP835.1 ACP869.1 ACP872.1 ACP874.1 ACP881.1 ACP883.1 ACT538.1 ACT548.1 ACT565.1 ACU349.1 ACU373.1 ACU381.1 ACU424.1 ADR679.1 ADR787.1 ADR794.1 ADR852.1 ADW762.1 ADW766.1 ADW771.1 AEH704.1 AEH733.1 AEH745.1 AEH748.1 AEH761.1 AEH764.1 AEH772.1 AEH826.1 AEH831.1 AEH856.1 AEH862.1 AEH865.1 AEH883.1 AEH951.1 AEH953.1 AEH955.1 AEH962.1 AEH972.1 AEJ054.1 AEJ100.1 AEJ101.1 AEJ106.1 AEJ111.1 AEJ145.1 AEJ148.1 AEJ164.1 AEJ167.1 AEJ203.1 AEJ241.1 AEJ251.1 AEJ252.1 AEJ257.1 AEM803.1 AEM833.1 AEM852.1 AEM857.1 AEM877.1 AEM885.1 AEM887.1 AEM888.1 AEM917.1 AEM951.1 AEM964.1 AEM967.1 AEM969.1 AEM988.1 AFB225.1 AFB349.1 AFB376.1 AFB425.1 AFB433.1 AFB434.1 AFB436.1 AFB444.1 AFB454.1 AFB495.1 AFB658.1 AFG344.1 AFG446.1 AFG471.1 AFG496.1 AFG545.1 AFG668.1 AFG672.1 AFG679.1 AFG704.1 AFG724.1 AFG947.1 AFG959.1 AFZ174.1 AFZ211.1 AFZ224.1 AFZ260.1 AFZ359.1 AFZ364.1 AFZ384.1 AFZ409.1 AFZ427.1 AFZ532.1 AFZ588.1 AFZ590.1 AFZ807.1 AFZ835.1 AFZ851.1 AFZ996.1 AGM602.1 AGM732.1 AGM765.1 AGM950.1 AGN231.1 AGN324.1 AGN447.1 AGN490.1 AGN552.1 AGN616.1 AHD572.1 AHD578.1 AHD589.1 AHD612.1 AHD646.1 AHD648.1 AHD663.1 AHD677.1 AHD686.1 AHD713.1 AHD818.1 AHD821.1 AHD822.1 AHK567.1 AHK727.1 AHK778.1 AHK897.1 AHK906.1 AHK922.1 AHK976.1 AHL049.1 AHL054.1 AHL069.1 AHL093.1 AHL095.1 AHL097.1 AHL105.1 AHL122.1 AHL133.1 AHL135.1 AHL138.1 AHL195.1 AHV557.1 AHV591.1 AHV603.1 AHV679.1 AHV705.1 AHV797.1 AHV801.1 AHV802.1 AHV807.1 AHV810.1 AHV818.1 AHV824.1 AHV843.1 AHV851.1 AHV854.1 AHV855.1 AHV860.1 AHV873.1 AHV921.1 AHV937.1 AHV945.1 AHV947.1 AHV965.1 AHV971.1 AHV975.1 AHV992.1 AHW028.1 AHW049.1 AHW055.1 AHW080.1 AHW082.1 AHW085.1 AHW104.1 AHW156.1 AHW489.1 AHW492.1 AHW496.1 AHW498.1 AHW661.1 AHW663.1 AHW670.1 AHW683.1 AJP994.1 AJP995.1 AJR053.1 AJR058.1 AJR073.1 AJR075.1 AJR076.1 AJR077.1 AJW280.1 AJW285.1 AJW287.1 AJW290.1 AJW296.1 AJW310.1 AJW328.1 AJW329.1 AJW334.1 AJW336.1 AJW345.1 AJW365.1 AJW382.1 AJW384.1 AJW388.1 AJW393.1 AJW412.1 AJW422.1 AJW431.1 AJW461.1 AJW487.1 AKX628.1 ALE668.1 ALE944.1 ALJ257.1 ALJ402.1 ALJ928.1 ALJ984.1 ALJ992.1 ALN367.1 ALN378.1 ALN429.1 ALN438.1 ALN441.1 ALN444.1 ALN448.1 ALN451.1 ALN469.1 ALN477.1 ALN487.1 ALN547.1 ALN593.1 ALN596.1 ALN622.1 ALN628.1 ALN634.1 ALN658.1 ALX037.1 ALX043.1 ALX053.1 ALX085.1 ALX203.1 ALX313.1 ALY252.1 ALY270.1 ALY272.1 ALY313.1 ALY329.1 ALY367.1 ALY369.1 ALY371.1 ALY400.1 ALY431.1 ALY533.1 ALY544.1 ALY610.1 ALY624.1 ALY625.1 ALY636.1 ALY653.1 ALY660.1 ALY774.1 ALY778.1 ALY799.1 ALY801.1 ALY802.1 ALY810.1 ALZ039.1 ALZ051.1 ALZ064.1 ALZ080.1 ALZ311.1 AMK053.1 AMN786.1 AMN963.1 AMP153.1 AMP226.1 AMS002.1 AMS004.1 AMS007.1 AMS022.1 AMS024.1 AMS026.1 AMS038.1 AMS068.1 AMS073.1 AMS126.1 AMS127.1 AMS161.1 AMS169.1 AMS193.1 AMS228.1 AMS241.1 AMS248.1 AMS256.1 AMS258.1 AMS259.1 AMS270.1 AMS271.1 AMS400.1 AMS431.1 AMS879.1 AMT051.1 AMT260.1 AMT351.1 ANG969.1 ANG990.1 ANH036.1 ANH126.1 ANH170.1 ANH185.1 ANH258.1 ANH278.1 ANH468.1 ANH503.1 ANH603.1 ANH618.1 ANH636.1 ANL856.1 YY2921.1 YY5780.1 YY5876.1 YY6049.1 YY6118.1 YZ1385.1 YZ1487.1 YZ1509.1 YZ1511.1 YZ1514.1 YZ1515.1 YZ1531.1 YZ1574.1 YZ1595.1 YZ9624.1 YZ9636.1 YZ9648.1 YZ9651.1 YZ9654.1 YZ9683.1 YZ9691.1 YZ9702.1 YZ9709.1 YZ9712.1 YZ9713.1 YZ9724.1 YZ9732.1 YZ9749.1 YZ9755.1 YZ9768.1 YZ9826.1 YZ9857.1 ZK3232.1 ZK3240.1 ZK3253.1 ZK3370.1 ZV3919.1 ZW0254.1 ZW0302.1 ZW0587.1 ZW8865.1 ZW8885.1 ZW8943.1 ZW8991.1 ZY8811.1 ZY8821.1 ZY9321.1 ZY9369.1 ZY9406.1

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.