Skip to content
Recall Observatory FDA recall evidence

Device product

RR Mechatronics Starrsed Control Level N ESR Control 4.6 ml QRR049001- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures. Product Code: A0026577

Z-1694-2023

April 14, 2023

Class III

Product summary

Firm
Mechatronics USA
Event
Event 92276
Status
Ongoing
Classification
Class III
Quantity
9306 units
Official record key
device-enforcement:Z-1694-2023

Official wording

Reason: QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.

Code information: UDI-DI: (01)08719189137118(17)240222(10)QCAB6AN505 Lot Number/Exp Date: QCAB6AN505 24-02-2022

Distribution pattern: Worldwide distribution - US Nationwide distribution in the state of OH and the country of Netherlands.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.