Device product

Philips EPIQ Diagnostic Ultrasound System; Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C, EPIQ CVx & EPIQ CVxi. MOD EPIQ Elite DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only 100-240V~50/60 Hz, 600 VA MOD EPIQ 5 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD EPIQ DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only,100-240V~50/60 Hz, 600VA

Z-0944-2021

December 16, 2020

Class II

Product summary

Firm
Philips Ultrasound Inc
Event
Event 87012
Status
Terminated
Classification
Class II
Quantity
607 units
Official record key
device-enforcement:Z-0944-2021

Official wording

Reason: The manufacturer discovered an issue associated with the EPIQ Image Boost with xPlane and Color Flow or Doppler while using the X8-2t TEE Transducer. If Image Boost is enabled the color box shows flow visualization not accurately represent the fluid flow. Similarly, if Image Boost is enabled, both CW and PW Doppler traces will not accurately represent the fluid flow.

Code information: All EPIQ Systems with software version 7.0 Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C, EPIQ CVx & EPIQ CVxi

Distribution pattern: US - AL, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE,NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI and WV OUS- Austria, Australia, Belgium, Canada, Chezh Republic, Croatia, Denmark, Egypt, Finland, France, Hong Kong, Indonesia, India, Ireland, Italy, Japan, Latvia, Lithuania, Malta, Mexico, Netherlands, Norway, Poland, Romania, Russia, Senegal, Slovenia, Sri Lanka, South Africa, Switzerland, Thailand, Turkey, UAE and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The manufacturer discovered an issue associated with the EPIQ Image Boost with xPlane and Color Flow or Doppler while using the X8-2t TEE Transducer. If Image Boost is enabled the color box shows flow visualization not accurately represent the fluid flow. Similarly, if Image Boost is enabled, both CW and PW Doppler traces will not accurately represent the fluid flow.