Skip to content
Recall Observatory FDA recall evidence

Device product

Integris Allura 9, system codes 722018 722021

Z-1488-2020

December 30, 2019

Class II

Product summary

Firm
Philips North America, LLC
Event
Event 84756
Status
Terminated
Classification
Class II
Quantity
20
Official record key
device-enforcement:Z-1488-2020

Official wording

Reason: A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

Code information: Serial Codes 8 114 64 906187 41 38 119 75 25 87 67 107 32 97

Distribution pattern: Nationwide domestic distribution. Foreign distribution worldwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.