Device product

Fresenius 2008K Series : Hemodialysis System

Z-1027-2017

December 21, 2016

Class II

Product summary

Firm
Fresenius Medical Care Renal Therapies Group, LLC
Event
Event 76037
Status
Terminated
Classification
Class II
Quantity
78481 (US) 2734 units (OUS)
Official record key
device-enforcement:Z-1027-2017

Official wording

Reason: When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

Code information: All serial numbers

Distribution pattern: Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine