Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 10mm x15mm (OUS) UPN: M00553550 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.

Z-0777-2021

December 01, 2020

Class II

Product summary

Firm
Boston Scientific Corporation
Event
Event 86909
Status
Terminated
Classification
Class II
Quantity
61 units
Official record key
device-enforcement:Z-0777-2021

Official wording

Reason: The rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity

Code information: Lot Number: 26156059, 26167275, 26167276, 26156742, 26156740 Exp Date: 10/9/2022 10/12/2022 GTIN: 08714729904571

Distribution pattern: US Nationwide distribution - Foreign: TBD.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity