Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

Z-2529-2023

July 26, 2023

Class I

Product summary

Firm
Hamilton Medical, Inc.
Event
Event 92873
Status
Ongoing
Classification
Class I
Quantity
6 units
Official record key
device-enforcement:Z-2529-2023

Official wording

Reason: Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Code information: UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595

Distribution pattern: Distribution US nationwide, including Puerto Rico. OUS: Not provided

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.