Device product
Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging.
Z-2315-2012
Product summary
- Event
- Event 62928
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3 units
- Official record key
device-enforcement:Z-2315-2012
Official wording
Reason: The firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil.
Code information: MRT-1503/P5-Vantage MRI System. Serial Codes: P5A0632002, P5B06Y2005, P5B0712006, P3F11Z2143.
Distribution pattern: US (nationwide) distribution including the states of Ohio and Texas.
Derived failure modes
-
Unknown
The firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil.