Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R) system with the heterogeneous module

Z-0561-2014

May 06, 2013

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 65131
Status
Terminated
Classification
Class II
Quantity
17,719
Official record key
device-enforcement:Z-0561-2014

Official wording

Reason: Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461. These lots shipped between 12/12/2012 and 3/22/2013. Internal investigation has confirmed shifts can be encountered in two scenarios: 1 - When an affected l

Code information: Lot numbers RD23031, exp 2013-04-29; RD23111, exp 2013-05-07; RD23241, exp 2013-05-19; RD23311, exp 2013-06-04; and RD23461, exp 2013-06-11

Distribution pattern: US Nationwide and OUS.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461. These lots shipped between 12/12/2012 and 3/22/2013. Internal investigation has confirmed shifts can be encountered in two scenarios: 1 - When an affected l