Skip to content
Recall Observatory FDA recall evidence

Device product

ARTISTE with syngo RT Therapist-linear accelerator systems to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer. Model Number: 8162815

Z-0441-2022

December 09, 2021

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 89228
Status
Terminated
Classification
Class II
Quantity
13 units
Official record key
device-enforcement:Z-0441-2022

Official wording

Reason: Potential safety issue under specific preconditons that may result in a user selecting the wrong site for treatment with the possibility to deliver dose to the wrong isocenter which could result in serious patient injury

Code information: Serial Numbers: 10085 10524 10737 10736 10586 10792 10793 10822 10843 10850 10886 10865 11122 UDI: 040568691613655911

Distribution pattern: US Nationwide distribution in the states of IA, LA, NE, NJ, NV, NY, OH, PA, SC, SD.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential safety issue under specific preconditons that may result in a user selecting the wrong site for treatment with the possibility to deliver dose to the wrong isocenter which could result in serious patient injury