Device product
APLIO i-series: a) APLIO i700 TUS AI700 b) APLIO i800 TUS AI800 diagnostic ultrasound system
Z-2712-2017
Product summary
- Event
- Event 77599
- Status
- Terminated
- Classification
- Class II
- Quantity
- 28 units
- Official record key
device-enforcement:Z-2712-2017
Official wording
Reason: When Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below.
Code information: Serial Numbers: a) JHA16X2002, HAB16Z2001, HAB16Z2002, HAB16Z2003, HAB16Z2004, HAB16Z2005, HAB1732007, HAB1732009 b) HAB16Z2002, HAB16Z2003, HAB1712006, HAB1712007, HAB1712008, HAB1712009, HAB1712010, HAB1722011, HAB1722012, HAB1732013, HAB1732014, HAB1732015, HAB1732016, HAB1732017, HAB1732018, HAB1732020, HAB1732021, HAB1732025, HAB1732027, HAB1732030
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown