Device product
DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724
Z-1677-2023
Product summary
- Event
- Event 92210
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 805 units
- Official record key
device-enforcement:Z-1677-2023
Official wording
Reason: IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Code information: UDI-DI:10886705002108 All lot numbers
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
Derived failure modes
-
Unknown
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time