Device product
Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A; Invue Lid: SI70095A; S-Lift Lid: SI50070A; S-Lift Base: SI50067A; Arena-C Lid: 11-72013D, 11-72014D; Generics Lid: 12-99005A. System cases are designed to hold and secure all instrumentation to be used with a surgical system.
Z-1958-2013
Product summary
- Event
- Event 64970
- Status
- Terminated
- Classification
- Class II
- Quantity
- 258 units (30 in distribution at the time of the recall)
- Official record key
device-enforcement:Z-1958-2013
Official wording
Reason: Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles.
Code information: Inspan Lid: Part Number: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A, Lot Numbers: 7197.024, 011649, 011498; Invue Lid: SI70095A, Lot Number: 6250.009-R; S-Lift Lid: SI50070A, Lot Number: 7201.008; S-Lift Base: SI50067A; Lot Numbers: 7183.007R, 7201.007; Arena-C Lid: 11-72013D, 11-72014D, Lot Numbers: 011969-001, 011969-002; Generics Lid: 12-99005A, Lot Number: 8372.001.
Distribution pattern: Worldwide Distribution-USA including the states of FL, OH, MA, and TX, and the countries of Mexico, and Dominican Republic.
Derived failure modes
-
Unknown
Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles.