Skip to content
Recall Observatory FDA recall evidence

Device product

REF 00-4309-071-40 Trabecular Metal Reverse Shoulder System Instrumentation, Glenosphere Helmet, 40 mm Diameter Nonsterile, Qty -1 Instrumentation used in shoulder replacement surgery. The glenosphere helmet is used to insert the glenosphere onto the base plate prior to impaction with the glenospere impactor.

Z-1954-2013

June 28, 2013

Class II

Product summary

Firm
Zimmer, Inc.
Event
Event 65502
Status
Terminated
Classification
Class II
Quantity
3,163 both sizes
Official record key
device-enforcement:Z-1954-2013

Official wording

Reason: There is potential for one or both tabsof theTrabecular Metal Reverse Glenosphere Helmet to fracture during use and become disassociated from the device. There have been 47 reported complaints of tab fracture.

Code information: 60720060 60729529 60782351 60875299 60902349 60948238 60943023 60985367 61019824 61034822 61063216 61145917 61183670 61207699 61218854 61356595 61400936 61515555 61566478 61615548 61756075 61780197 61832594 61864229 61923988 61978214 62041752 62136099 62180444 62226761 and 62289420

Distribution pattern: Worldwide Distribution - USA (nationwide) and internationally to Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Israel, Italy, Korea, Lebanon, Malta, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom, Utd.Arab.Emir.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is potential for one or both tabsof theTrabecular Metal Reverse Glenosphere Helmet to fracture during use and become disassociated from the device. There have been 47 reported complaints of tab fracture.