Skip to content
Recall Observatory FDA recall evidence

Device product

The Sarns TCM II (system) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. It also freezes water for an ice supply, monitors temperatures in the patient and extracorporeal circuit, and allows gradient rewarming relative to a venous blood temperature. The Sarns" TCM II also features a Cardioplegia System which will supply cooling water for cardioplegia. Device Name / Model Number: TCM II, with Cardioplegia, 110V/60Hz (P/N 164925) TCM II, without Cardioplegia, 110V/60Hz (P/N 164930) TCM II, with Cardioplegia, 220V/50Hz (P/N 164935) TCM II, without Cardioplegia, 220V/50Hz (P/N 164940) TCM (P/N 15747) Catalog Number: 4415, 164925, 164930, 164935, 15747

Z-1791-2021

April 16, 2021

Class II

Product summary

Firm
Terumo Cardiovascular Systems Corporation
Event
Event 87834
Status
Terminated
Classification
Class II
Quantity
995 devices
Official record key
device-enforcement:Z-1791-2021

Official wording

Reason: Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

Code information: All lot numbers distributed from 05/02/1985 thru 06/10/2015

Distribution pattern: Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.