Skip to content
Recall Observatory FDA recall evidence

Device product

CRTD DTPA2D1G COBALT XT HF QUAD OUS, Model Number DTPA2D1G; Implantable Cardioverter Defibrillators

Z-1775-2023

May 10, 2023

Class I

Product summary

Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Event
Event 92328
Status
Ongoing
Classification
Class I
Quantity
5 units
Official record key
device-enforcement:Z-1775-2023

Official wording

Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Code information: GTIN 00763000501044, lot serial Numbers: RTH602452S, RTH602446S, RTH602443S, RTH602454S, RTH602442S

Distribution pattern: US Nationwide - Worldwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.