Skip to content
Recall Observatory FDA recall evidence

Device product

The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right). The system has the capacity to circulate water at a rate of up to 6.5 gal./min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Device Name / Model Number: HX2 Temperature Management System (P/N 809810) Catalog Number: 809810

Z-1789-2021

April 16, 2021

Class II

Product summary

Firm
Terumo Cardiovascular Systems Corporation
Event
Event 87834
Status
Terminated
Classification
Class II
Quantity
75 devices
Official record key
device-enforcement:Z-1789-2021

Official wording

Reason: Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

Code information: All lot numbers distributed from 05/02/1985 thru 06/10/2015

Distribution pattern: Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.