Skip to content
Recall Observatory FDA recall evidence

Device product

VITROS Chemistry Products CKMB Slides: (1)VITROS CKMB Slides (60 Slide cartridge) Product Code:8058232 (2)VITROS CKMB Slides (18 Slide cartridge) Product Code: 8001133 Quantitatively measure creatine kinase MB (CK-MB) activity in serum using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems.

Z-0770-2021

December 11, 2020

Class II

Product summary

Firm
Ortho Clinical Diagnostics Inc
Event
Event 87028
Status
Ongoing
Classification
Class II
Quantity
Total= 41,598 U.S. = 403 xU.S. = 41,195; Extended: 25,775 units
Official record key
device-enforcement:Z-0770-2021

Official wording

Reason: VITROS Chemistry Products CKMB Slides Potential for Falsely Elevated Results Due to CK-MM Interference

Code information: INITIAL Communication: 11 December 2020, (Communication Ref. # CL2020-302) Coatings 0230, 0231, 0233, 0235, 0239 (all lot numbers) Extended Recalls: Recall Extension #1, for VITROS CKMB Slides, Coating 0245 (Communication Reference, CL2021-182) 18 June 2021 Recall Extension #2 for VITROS CKMB Slides, Coating 0246 (Communication Reference, CL2021-214) 18 August 2021 Recall Extension #3 for VITROS CKMB Slides, Coating 0248 (Communication Reference, CL2021-274) , 1 November 2021 Recall Extension #4 for VITROS CKMB Slides, Coating 0261 (Communication Reference, CL2023-244) , 27 October 2023 UDI: (1) (10758750004294) (2) (10758750004201

Distribution pattern: Worldwide distribution - US Nationwide distribution and the countries of Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore 609917, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    VITROS Chemistry Products CKMB Slides Potential for Falsely Elevated Results Due to CK-MM Interference