Artis Q.zen floor, Model Number 10848353

Z-0978-2022

February 23, 2022

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 89860
Status
Ongoing
Classification
Class II
Quantity
33 systems
Official record key
device-enforcement:Z-0978-2022

Official wording

Reason: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Code information: UDI: 4056869010014; Serial Numbers: 105120 105110 105033 105032 105500 105105 105028 105503 105018 105114 105115 105026 105025 105135 105016 105119 105137 105501 105132 105128 105142 105017 105121 105122 105123 105100 105112 105124 105029 105030 105031 105125 Additional Serial Numbers as of 10/4/23: 143500

Distribution pattern: US Nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".