Device product
Critical Care Ventilator, Catalog Number(s): 8422300: Evita V600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.
Z-1690-2021
Product summary
- Event
- Event 87819
- Status
- Terminated
- Classification
- Class II
- Quantity
- 34 units
- Official record key
device-enforcement:Z-1690-2021
Official wording
Reason: Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
Code information: All lots/serial numbers manufactured from 1/1/2019-3/31201
Distribution pattern: US Nationwide distribution in the states of PA, TX, MD, FL, TX, WA, MS, NY.
Derived failure modes
-
Unknown
Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning