Skip to content
Recall Observatory FDA recall evidence

Device product

Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX and ComboLab AltiX.

Z-2613-2023

August 08, 2023

Class II

Product summary

Firm
GE Medical Systems Information Technologies Inc
Event
Event 92903
Status
Ongoing
Classification
Class II
Quantity
81 amplifiers
Official record key
device-enforcement:Z-2613-2023

Official wording

Reason: A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.

Code information: GTIN 00195278507044 (CardioLab) and 00195278507051 (ComboLab) Serial numbers for Amplifiers: SVS22330007SA, SVS22330008SA, SVS22330009SA, SVS22330010SA, SVS22330011SA, SVS22330012SA, SVS22330013SA, SVS22330014SA, SVS22330016SA, SVS22330017SA, SVS22330018SA, SVS22330019SA, SVS22330020SA, SVS22330021SA, SVS22330022SA, SVS22330023SA, SVS22330024SA, SVS22330025SA, SVS22330026SA, SVS22330028SA, SVS22330030SA, SVS22330031SA, SVS22330032SA, SVS22330033SA, SVS22330034SA, SVS22330035SA, SVS22440001SA, SVS22440002SA, SVS22440005SA, SVS22440008SA, SVS22440009SA, SVS22440010SA, SVS22440011SA, SVS22440012SA, SVS22440013SA, SVS22440016SA, SVS22490002SA, SVS22490003SA, SVS22490004SA, SVS22520006SA, SVS22520008SA, SVS22520009SA, SVS22520010SA, SVS22520011SA, SVS22520012SA, SVS22520013SA, SVS22520015SA, SVS23040001SA, SVS23040002SA, SVS23040003SA, SVS23040004SA, SVS23040005SA, SVS23040007SA, SVS23040008SA, SVS23040011SA, SVS23040012SA, SVS23040013SA, SVS23040014SA, SVS23040015SA, SVS23040016SA, SVS23040017SA, SVS23040018SA, SVS23040019SA, SVS23070002SA, SVS23070004SA, SVS23070005SA, SVS23070008SA, SVS23070009SA, SVS23070011SA, SVS23070012SA, SVS23100001SA, SVS23100002SA, SVS23100004SA, and SVS23100006SA. Serial numbers for FRUs: SVS22330015SA, SVS22440006SA, SVS22440007SA, SVS22440015SA, SVS22490001SA, SVS22520005SA, and SVS22520014SA.

Distribution pattern: Distribution was made to CA, CO, FL, GA, IL, IN, KY, MA, NC, NJ, OH, OR, PA, SC, TX, VA, and WI. There was government distribution but no military distribution. Foreign distribution was made to Canada, France, Germany, Greece, Ireland, Japan, Republic of Korea, and Romania.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.