Skip to content
Recall Observatory FDA recall evidence

Device product

Implant Retrieval Instrument Hex & Tri-Ch NP/RP 22mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.

Z-1316-2020

March 14, 2019

Class II

Product summary

Firm
Nobel Biocare Usa Llc
Event
Event 84450
Status
Terminated
Classification
Class II
Quantity
2470 units
Official record key
device-enforcement:Z-1316-2020

Official wording

Reason: Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.

Code information: (01)07332747083767(10)73491 (01)07332747083767(10)73600 (01)07332747083767(10)73296 (01)07332747083767(10)73315 (01)07332747083767(10)73595 Catalog# 37471

Distribution pattern: Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.