Sharesource Connectivity Platform for Use with the Amia Automated PD System

Z-2173-2021

June 30, 2021

Class II

Product summary

Firm
BAXTER HEALTHCARE CORPORATION
Event
Event 88219
Status
Terminated
Classification
Class II
Quantity
1 unit
Official record key
device-enforcement:Z-2173-2021

Official wording

Reason: Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.

Code information: Product Code 5C9400, UDI: 45413765565692, 55413765565699

Distribution pattern: Worldwide distribution - US Nationwide distribution and the country of Japan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.