Device product
Kaguya Automated Peritoneal Dialysis System
Z-2174-2021
Product summary
- Event
- Event 88219
- Status
- Terminated
- Classification
- Class II
- Quantity
- 5101 units
- Official record key
device-enforcement:Z-2174-2021
Official wording
Reason: Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.
Code information: Product Code T5C8500 (distribution in Japan only)
Distribution pattern: Worldwide distribution - US Nationwide distribution and the country of Japan.
Derived failure modes
-
Unknown
Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.