Skip to content
Recall Observatory FDA recall evidence

Device product

GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope System Per PLRA: The Core OneTouch Cables (Reusable Device) is ancillary equipment used with the Core 10 and Core 15 monitors. The monitor and workstation are intended to work with video endoscopes, in conjunction with ancillary equipment, for endoscopic procedures.

Z-2476-2020

June 05, 2020

Class I

Product summary

Firm
Verathon, Inc.
Event
Event 85705
Status
Terminated
Classification
Class I
Quantity
437
Official record key
device-enforcement:Z-2476-2020

Official wording

Reason: When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.

Code information: All serial numbers

Distribution pattern: Worldwide Distribution: US (nationwide) and countries of: Australia and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.