BD MAX SARS-CoV-2 Reagents, Cat. No. 445003 (UDI 00382904450038) and 445003-01 (UDI (01)603829044500301) - Product Usage: intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.

Z-2754-2020

July 20, 2020

Class II

Product summary

Firm
Becton Dickinson & Co.
Event
Event 86061
Status
Terminated
Classification
Class II
Quantity
30,392 to date
Official record key
device-enforcement:Z-2754-2020

Official wording

Reason: BD is experiencing elevated rates of potential false positive results from certain customers when using the BD SARS-CoV-2 Reagents for the BD MAX System. False positive test results could lead to a delay in diagnosis and potentially cause the patient to be exposed to COVID-19 if they are isolated with an infected individual.

Code information: All lots are affected

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AZ, CA, CO, DC, FL, GU, HI, IL, IN, KY, MD, MI, MO, NC, NJ, NY, PA, SC, TN, TX, WI, WY and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, Hong Kong, Japan, Mexico, New Zealand, Peru, Philippines, Singapore.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    BD is experiencing elevated rates of potential false positive results from certain customers when using the BD SARS-CoV-2 Reagents for the BD MAX System. False positive test results could lead to a delay in diagnosis and potentially cause the patient to be exposed to COVID-19 if they are isolated with an infected individual.