Artis icono floor with software VE20C-diagnostic imaging angiography system Model: 11327700

Z-0517-2022

October 12, 2021

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 89260
Status
Terminated
Classification
Class II
Quantity
23 units
Official record key
device-enforcement:Z-0517-2022

Official wording

Reason: Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option

Code information: Serial Numbers: 170319 170026 170343 170305 170307 170331 170309 170310 170023 170304 170328 170334 170336 170025 170346 170051 170333 170332 170312 170027 170303 170322 170053 UDI: 04056869149325

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option