Skip to content
Recall Observatory FDA recall evidence

Device product

Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps.

Z-2348-2012

August 12, 2009

Class II

Product summary

Firm
Terumo Cardiovascular Systems Corporation
Event
Event 62321
Status
Terminated
Classification
Class II
Quantity
8
Official record key
device-enforcement:Z-2348-2012

Official wording

Reason: Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav

Code information: catalog number 16404 and serial numbers: 4357, 4381, 4386, 4387, 4389, 4391, and 4392.

Distribution pattern: Worldwide distribution: International only-country of: Terumo Dubai, UAE

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav