AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Z-1364-2022

August 03, 2021

Class II

Product summary

Firm
Fresenius Medical Care Holdings, Inc.
Event
Event 90411
Status
Terminated
Classification
Class II
Quantity
2 units
Official record key
device-enforcement:Z-1364-2022

Official wording

Reason: AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

Code information: UDI-DI: (01)04251285200227, (01)04251285200289, (01)0425128520; Serial Numbers: 7BPS1621 7BDS0783 7BFS0517 7BPS1611 7BDS0776 7BFS0502

Distribution pattern: US Nationwide distribution in the states of NC and WV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation