Skip to content
Recall Observatory FDA recall evidence

Device product

TX1 Tissue Removal System Console. Model: TX1 TX1 Tissue Removal System is an ultrasonic aspirator that emulsifies and removes soft tissue. The TX1 Console provides control over each mode of operation, as well as the user interface via LCD touchscreen and audible tones for confirmation of user selections. Indiciated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery.

Z-1817-2012

February 22, 2012

Class II

Product summary

Firm
American Optisurgical Inc
Event
Event 61253
Status
Terminated
Classification
Class II
Quantity
5units
Official record key
device-enforcement:Z-1817-2012

Official wording

Reason: Two reasons for recall. 1. Incorrect pressure relief valve installed inside console, leading to a remote chance of a risk to health. 2. Technical Specifications, under Irrigation, the Operating Pressure value is incorrect.

Code information: Product Codes: LFL, GEI, Serial #: 55400001, 55400002, 55400003. Part Order No: 554-0034-000 510(K) Mp.: K101561

Distribution pattern: Nationwide distribution: MN only.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Two reasons for recall. 1. Incorrect pressure relief valve installed inside console, leading to a remote chance of a risk to health. 2. Technical Specifications, under Irrigation, the Operating Pressure value is incorrect.