Device product
Mini-Infuser Model 300XL Multispeed Infusion Pump, an Rx battery operated syringe infusion pump with an active time scale and rate selector switch; product codes 2M8171 and 2M8171R. Product Usage: These product codes includes refurbished units. Intended for the controlled rate delivery of small volume parental fluids as prescribed by a physician. Usage: The Mini-Infuser System is designed for controlled intermittent administration of intravenous medication.
Z-2269-2012
Product summary
- Event
- Event 62881
- Status
- Terminated
- Classification
- Class III
- Quantity
- 19,380 units
- Official record key
device-enforcement:Z-2269-2012
Official wording
Reason: Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum
Code information: product codes 2M8171 and 2M8171R, all serial numbers
Distribution pattern: 45,865 units
Derived failure modes
-
Unknown
Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum