Skip to content
Recall Observatory FDA recall evidence

Device product

IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only.

Z-2124-2012

July 09, 2012

Class III

Product summary

Firm
Iris Diagnostics
Event
Event 62555
Status
Terminated
Classification
Class III
Quantity
862
Official record key
device-enforcement:Z-2124-2012

Official wording

Reason: The firm recalled after it received complaints of IRISpec CA/CB two part Urine Chemistry Control, 101131-12, where the nitrite results read negative for the part B control, which should be positive.

Code information: Iris Diagnostics part number 800-0074, lot number 131-12.

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm recalled after it received complaints of IRISpec CA/CB two part Urine Chemistry Control, 101131-12, where the nitrite results read negative for the part B control, which should be positive.