GE Healthcare, Aestiva/5 7900 SmartVent, anesthesia machine. Used in Datex-Ohmeda Aestiva/5 Anesthesia Systems.

Z-1787-2012

April 10, 2012

Class I

Product summary

Firm
GE Healthcare, LLC
Event
Event 61639
Status
Terminated
Classification
Class I
Quantity
8
Official record key
device-enforcement:Z-1787-2012

Official wording

Reason: GE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent at the same time. This could manifest as over delivery of a single agent if both vaporizers contain the same agent or as delivery of more than one agent. This m

Code information: Serial Number AMRP01031 AMRP00966 AMRP01030 AMRP00968 AMRP00967 AMRP01033 AMRP00970 AMRP00969

Distribution pattern: Nationwide Distribution - including the states of AL, FL, GA, MO, and VA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    GE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent at the same time. This could manifest as over delivery of a single agent if both vaporizers contain the same agent or as delivery of more than one agent. This m