"***Electric Systems Foot Control with Direction Only***Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium.
Z-0075-2013
Product summary
- Event
- Event 63086
- Status
- Terminated
- Classification
- Class II
- Quantity
- 41 units
- Official record key
device-enforcement:Z-0075-2013
Official wording
Reason: The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body.
Code information: Serial Numbers: E28305271001, E28305271002, E28305271003, E28305271004, E2830527100, E28305271006, E28305271007, E28305271008, E2830521009, E28305271010, E29305270901, E29305270902, E29305270903, E29305271101, E29305271102, E29305271103, E29305271104, E29305271105, E29305271106, E29305271107, E29,
Distribution pattern: Worldwide Distribution - US (nationwide) and the countries of: Italy, Japan, Switzerland, Turkey and Taiwan.
Derived failure modes
-
Unknown
The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body.