Terumo Advanced Perfusion System 1 Flow Module indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Z-2449-2012

September 20, 2012

Class II

Product summary

Firm
Terumo Cardiovascular Systems Corporation
Event
Event 62322
Status
Terminated
Classification
Class II
Quantity
72 units
Official record key
device-enforcement:Z-2449-2012

Official wording

Reason: Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on the flow module was not illuminated. Terumo CVS' investigation found that the malfunction was caused by an improper component on a circuit board used in a populat

Code information: Catalog number: 802018 and serial numbers: 00122,00132, 00137, 00321-00337, 00339, 00340, 00342-00389,00391-00439, 00441, 00442, 00444-00455, 00457, 00459, 00460, 00466, 00467, 00475, and 00488.

Distribution pattern: Worldwide Distribution - USA (nationwide) and the countries of AUSTRALIA, Singapore, Jordan, Japan, Vietnam, Germany, Russia, Saudi Arabia, BELGIUM, Mexico, and CANADA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on the flow module was not illuminated. Terumo CVS' investigation found that the malfunction was caused by an improper component on a circuit board used in a populat