Unicel Closed Tube Aliquotter (UCTA) for UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, UniCel DxC 880i Synchron Access Clinical Systems. Part Number A45853. Product Usage: The UniCel DxC Synchron Clinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in-vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).

Z-0059-2013

July 06, 2012

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 63191
Status
Terminated
Classification
Class II
Quantity
18 units
Official record key
device-enforcement:Z-0059-2013

Official wording

Reason: The recall was initiated because Beckman Coulter has confirmed that the current sample syringe pump drive assemblies installed on DxC Integrated ystem with UniCel Closted Tube Aliquotter (UCTA) serial numbers between 547 and 574 may cause a hardware motion error (ID0x08000501).

Code information: Serial Numbers: 547-574 (UCTA)

Distribution pattern: Worldwide Distribution and US (nationwide) and the countries of China, France, Italy and Taiwan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The recall was initiated because Beckman Coulter has confirmed that the current sample syringe pump drive assemblies installed on DxC Integrated ystem with UniCel Closted Tube Aliquotter (UCTA) serial numbers between 547 and 574 may cause a hardware motion error (ID0x08000501).