Device product
Boston Scientific, Z Flex 270, Steerable Sheath, Sterile EO, Rx only, Distributed by Boston Scientific Corporation, One Boston Scientific Place, Natick, MA 01760-1537, Manufactured by Greathbatch Medical, 2300 Berkshire Lane North, Minneapolis, MN 55441. To facilitate transvenous introduction of diagnostic / therapeutic catheters into the vasculature and into the chambers of the heart.
Z-2459-2012
Product summary
- Event
- Event 63141
- Status
- Terminated
- Classification
- Class II
- Quantity
- 787
- Official record key
device-enforcement:Z-2459-2012
Official wording
Reason: During the investigation of four returned Z Flex 270 Steerable Sheaths that were used with cardiac cryoablation balloon catheters, the internal PTFE sheath liners were found to display varying degrees of damage that could pose an embolic risk to a patient.
Code information: Q2112752, Q2110939, Q2130137, Q2131250, Q2130212, Q2150143, W2162515, W2163769, W2160626, W2175891, W2172722, W2184492, W2196366, W2184490, W2205843
Distribution pattern: Distributed in MA.
Derived failure modes
-
Unknown