PathFinder NXT Pivoting Percutaneous Rod Holder This device is an instrument used during orthopedic surgery. The intended use of the pivoting percutaneous rod holder is to attach to the rod to facilitate rod insertion, positioning, and orientation through the extender sleeves, under the fascia. The rod is secured to the holder by tightening a captive set screw against the rod end in the tip of the holder using the set screw driver. After positioning the rod and securing at least one pedicle screw closure top against the implanted rod, the holder is removed from the rod end by loosening the captive set screw with the set screw driver.

Z-1933-2012

June 05, 2012

Class II

Product summary

Firm
Zimmer, Inc.
Event
Event 62137
Status
Terminated
Classification
Class II
Quantity
77 units
Official record key
device-enforcement:Z-1933-2012

Official wording

Reason: Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted rod from the rod holder instrument. The complaint investigation showed the hex mating feature of the set screw was broken and material missing. There were no reports of fragments of the set screw being left in the patient or that there was any patient or user injury, but these events we

Code information: Part number 3564-1; all lots

Distribution pattern: Worldwide distribution: USA (nationwide) and countries of: Australia, France and Germany.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted rod from the rod holder instrument. The complaint investigation showed the hex mating feature of the set screw was broken and material missing. There were no reports of fragments of the set screw being left in the patient or that there was any patient or user injury, but these events we