Device product
00780303601 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780303602 MIS 12/14 Slotted Femoral Head Provisional +0X36 00780303603 MIS 12/14 Slotted Femoral Head Provisional +3.5X36 00780303604 MIS 12/14 Slotted Femoral Head Provisional +7X36 00780303605 MIS 12/14 Slotted Femoral Head Provisional +10.5X36 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Z-2483-2012
Product summary
- Firm
- Zimmer, Inc.
- Event
- Event 62110
- Status
- Terminated
- Classification
- Class II
- Quantity
- 231,641 all lots, all sizes
- Official record key
device-enforcement:Z-2483-2012
Official wording
Reason: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
Code information: All lots including lot 78939000 (1/7/2002) to present
Distribution pattern: Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
Derived failure modes
-
Unknown
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp