Skip to content
Recall Observatory FDA recall evidence

Device product

Siemens Healthcare ADVIA Centaur CA 19-9 250 Test 10491244 50 Test 10491379 Intended Use The ADVIA Centaur CA 19-9 Assay is an in vitro immunoassay for the quantitative measurement of the CA 19-9 tumor-associated antigen, in human serum, using the ADVIA Centaur and ADVIA Centaur XP systems. This assay is indicated for the serial measurement of CA 19-9 to aid in the management of patients diagnosed with cancers of the exocrine pancreas.

Z-0508-2013

November 14, 2012

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc
Event
Event 63682
Status
Terminated
Classification
Class II
Quantity
250 test=525 and 50 test=12243
Official record key
device-enforcement:Z-0508-2013

Official wording

Reason: Operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Code information: All lot codes

Distribution pattern: Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.