Device product
Baxter Interlink System Buretrol Solution Set with 150 mL Burette (Ball-Valve Drip Chamber); a sterile Rx IV fluid pathway device. Product Usage: The device is used for the administration of fluids from a container into the patient's vascular system through a vascular access device.
Z-0012-2013
Product summary
- Event
- Event 63059
- Status
- Terminated
- Classification
- Class I
- Quantity
- 278,420 units
- Official record key
device-enforcement:Z-0012-2013
Official wording
Reason: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
Code information: product code 2C7519, all lots
Distribution pattern: Worldwide Distribution - US Nationwide and Internationally to Canada, Costa Rica, Australia, New Zealand, Hong Kong, and the United Kingdom.
Derived failure modes
-
Unknown
Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.