SenTec V-Sign Membrane Changer Insert (VS-MCI), 5 pieces. The SenTec V-Sign Membrane Changer Insert is indicated for use with the V-Sign Membrane Changer only. The V-sign Sensor is indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are required for adult and pediatric patients.

Z-0434-2013

November 05, 2012

Class II

Product summary

Firm
Sentec AG
Event
Event 63631
Status
Terminated
Classification
Class II
Quantity
265 boxes (5 per box)
Official record key
device-enforcement:Z-0434-2013

Official wording

Reason: A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.

Code information: Lot numbers: 1D0F, 1D58, 1D80, 1DCA, 1DEA, 1E5F

Distribution pattern: Distributed only in MO.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing defect