Skip to content
Recall Observatory FDA recall evidence

Device product

Stryker XIA 3 TORQUE WRENCH REF 48237028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Z-0380-2013

August 09, 2012

Class II

Product summary

Firm
Stryker Spine
Event
Event 63386
Status
Terminated
Classification
Class II
Quantity
54
Official record key
device-enforcement:Z-0380-2013

Official wording

Reason: Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Code information: Class I device, 510(K) exempt Catalog Number: 48237028 Lot Number: 093653, 093654, 093655

Distribution pattern: Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.