Sekisui Diagnostics, Spectrolyse PAI-1 In-Vitro Diagnostic Use for the quantitative determination of Plasminogen Activator Inhibitor Type 1 in human plasma. Catalog # 101201. Spectrolyse PAI-1 is intended for the quantitative determination of Plasminogen Activator Inhibitor Type-1 (PAI-1) activity in human plasma. The test is for in vitro diagnostic use.

Z-0463-2013

October 24, 2012

Class II

Product summary

Firm
Sekisui Diagnostics Llc
Event
Event 63615
Status
Terminated
Classification
Class II
Quantity
63 kits
Official record key
device-enforcement:Z-0463-2013

Official wording

Reason: Spectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutions.

Code information: Lot # 120827, exp 11/27/2012

Distribution pattern: Nationwide Distribution-including the states of FL, IN, NC, OH, UT, and WA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Spectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutions.