Skip to content
Recall Observatory FDA recall evidence

Device product

P10 Tobii battery box (product number 510410) Product Usage - Battery pack for MyTobii P10 Device. MyTobii P10 is a portable eye-controlled communication device featuring the Tobii Eye Control system.

Z-0766-2013

January 14, 2013

Class II

Product summary

Firm
Tobii Assistive Technology, Inc.
Event
Event 64099
Status
Terminated
Classification
Class II
Quantity
483 units
Official record key
device-enforcement:Z-0766-2013

Official wording

Reason: Tobii ATI customers P10 external battery pack, which contains sealed lead-acid batteries, malfunctioned while recharging and produced smoke with a chemical smell.

Code information: Product Code 510410. All units.

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Tobii ATI customers P10 external battery pack, which contains sealed lead-acid batteries, malfunctioned while recharging and produced smoke with a chemical smell.