Device product
Battery, 220V (Includes charger and connecting cable) The Sarns" Centrifugal System is indicated for use in cardiopulmonary bypass procedures only.
Z-0050-2013
Product summary
- Event
- Event 63079
- Status
- Terminated
- Classification
- Class II
- Quantity
- 289 units
- Official record key
device-enforcement:Z-0050-2013
Official wording
Reason: Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2004 of various malfunctions of the battery backup system for the Sarns" Centrifugal System. The reports can be broadly categorized as follows: " Failure to charge " Failure to hold charge " Failure to operate In the most severe case, a battery malfunction could result in an unexpected loss of arterial flow. In
Code information: Catalog number: 9491; and lot numbers: 1001-1187, 1101, 1193-1286, 2083-2085, 2018, 2288, 2289, 2447-2451, 2634, 2635, and 2884.
Distribution pattern: Worldwide distribution: USA including: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, UT, VA, WA, WI, and WV. Puerto Rico and internationally to: ALEXANDRIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CENTRO AMERICA, CHILE, COLOMBIA, Dominican Republic, Ecuador, FRANCE, GERMANY, GREECE, Guatamala, HOLLAND, HONG KONG, India, ITALY, Japan, KL SEGUNDO, KOREA, LEICESTERSHIRE, MALAYSA, Malaysia, Mexico, Mexico, Mexico, NETHERLANDS, Panama, Peru, Philippines, RALEIGH, Republic of Georgia, SANTIAGO CHILE, Saudi Arabia, SEOUL KOREA, SHANGHAI, CHINA, Singapore, South Africa, SPAIN, SWITZERLAND, TAIWAN, Thailand, UNITED ARAB EMIRATES (UAE), Uruguay, USA, and Venezuela,
Derived failure modes
-
Unknown
Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2004 of various malfunctions of the battery backup system for the Sarns" Centrifugal System. The reports can be broadly categorized as follows: " Failure to charge " Failure to hold charge " Failure to operate In the most severe case, a battery malfunction could result in an unexpected loss of arterial flow. In