Device product

Li-Ion battery used in the Oxus Portable Oxygen Concentrator (POC), RS-00400. Also branded as Reliability Plus byDrive. The Portable Oxygen Concentrator is intended to deliver concentrated oxygen for adult patients with chronic pulmonary diseases.

Z-0768-2013

January 14, 2013

Class II

Product summary

Firm
Oxus America, Inc.
Event
Event 64057
Status
Terminated
Classification
Class II
Quantity
1724 total units
Official record key
device-enforcement:Z-0768-2013

Official wording

Reason: Due to past battery inventory storage conditions and deficiencies in the maintenance procedure, it was possible that some batteries were over-discharged to point of damaging the cells.

Code information: part number RS-00402 Unit ID # (TS-): 001418, 001419, 001420, 001421, 001422, 001423, 001424, 001425, 001426, 001427, 001428, 001429, 001430, 001431, 001432, 001433, 001434, 001435, 001436, 001437, 001438, 001439, 001440, 001441, 001442, 001443, 001444, 001445, 001446, 001447, 001448, 001449, 001450, 001451, 001452, 001453, 001454, 001455, 001456, 001457, 001458, 001459, 001460, 001461, 001462, 001463, 001464, 001465, 001466, 001467, 001468, 001469, 001470, 001471, 001472, 001473, 001474, 001475, 001476, 001477, 001478, 001479, 001480, 001481, 001482, 001483, 001484, 001485, 001486, 001487, 001488, 001489, 001490, 001491, 001492, 001493, 001494, 001495, 001496, 001497, 001498, 001499, 001551, 001552, 001557, 001558, 001559, 001560, 001561, and 001562. Battery SN (MB-): 06625x, 06666x, 06679x, 06715x, 08335x, 06498x, 01578x, 12009x, 12010x, 00075x, 05624x, 11868x, 11855x, 11587x, 11601x, 08240x, 08336x, 06145x, 06160x, 06302x, 06305x, 08189x, 08190x, 09200x, 09562x, 09564x, 09654x, 09719x, 09720x, 09787x, 09788x, 09887x, 09902x, 09903x, 09904x, 09905x, 09906x, 09957x, 09958x, 09959x, 09960x, 09961x, 10049x, 10477x, 10478x, 10479x, 10480x, 05689x, 05690x, 05694x, 05695x, 05696x, 05727x, 07937x, 07938x, 07940x, 07941x, 08072x, 08112x, 08125x, 10358x, 11165x, 11166x, 11961x, 11027x, 08628x, 08630x, 09143x, 09144x, 09145x, 09146x, 09498x, 09888x, 09889x, 09890x, 09891x, 09983x, 09984x, 10547x, 10549x, 10550x, 01476x, 01527x, 01576x, 05630x, 08657x, 08677x, 08752x, 08753x, 08755x, 08838x, 09724x, 10237x, 10582x, 10597x, 10600x, 11252x, 11253x, 11264x, 11265x, 11266x, 11271x, 11272x, 11469x, 11472x, 11473x, 11525x, 11526x, 11527x, 11528x, 11999x, 01518x, 09212x, 10584x, 10602x, 11110x, 11426x, 11427x, 11456x, 11471x, 11478x, 11485x, 11486x, 11514x, 11518x, 11589x, 11591x, 11592x, 11593x, 05647x, 05648x, 05649x, 06646x, 06672x, 06673x, 06674x, 06675x, 06676x, 06702x, 07558x, 07637x, 07638x, 07639x, 07640x, 07641x, 07643x, 07644x, 07645x, 07663x, 07680x, 07689x, 07691x, 07697x, 07703x, 07704x, 07705x, 08257x, 08258x, 08259x, 08260x, 08649x, 08650x, 08681x, 08807x, 08808x, 08811x, 09304x, 09387x, 09388x, 09390x, 09859x, 09860x, 10089x, 10402x, 10613x, 10614x, 10616x, 05407x, 06072x, 06078x, 06080x, 06081x, 06209x, 06217x, 06226x, 06277x, 06278x, 06279x, 06299x, 06329x, 06331x, 06396x, 06432x, 06433x, 06434x, 06436x, 06442x, 06443x, 06444x, 06445x, 06446x, 06474x, 06475x, 06476x, 06481x, 06512x, 06515x, 06517x, 06518x, 06522x, 06525x, 06810x, 06872x, 06874x, 06877x, 07030x, 07112x, 07113x, 07114x, 07116x, 07154x, 07169x, 07170x, 07171x, 07173x, 07194x, 05589x, and 07749x.

Distribution pattern: Worldwide Distribution-USA including the states of New York, Michigan, Colorado, Texas, Iowa, Florida, Georgia, Ohio, Wisconsin, Oregon, Alabama, Illinois, Utah, Oklahoma, West Virginia and California and the countries of Canada, European Union, South Africa, Korea and Brazil.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to past battery inventory storage conditions and deficiencies in the maintenance procedure, it was possible that some batteries were over-discharged to point of damaging the cells.