Device product
HemoCue¿ Glucose 201 Microcuvettes, 25 Microcuvettes Individually Packaged. The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 Analyzer and the HemoCue Glucose 201 DM Analyzer. Quantitative determination of glucose in whole blood using a specially designed analyzer, the HemoCue Glucose 201 with Plasma conversion. The quantitative determination of the instant blood glucose concentration in circulation supplements the clinical evidence in the diagnosis and treatment of patients with diabetes as well as monitoring of neonatal blood glucose levels.
Z-2190-2013
Product summary
- Firm
- HemoCue AB
- Event
- Event 65854
- Status
- Terminated
- Classification
- Class II
- Quantity
- 22,215 boxes of 100 cuvettes distributed in US, 34, 841 distributed outside the US
- Official record key
device-enforcement:Z-2190-2013
Official wording
Reason: During a complaint investigation HemoCue AB discovered punctured HemoCue Glucose 201 single pack pouches.
Code information: Model #, 110705, Lot #, 1207767, 1208500; 1208501; 1208772; 1208773; 1208775; 1208777-1208781; 1208787; 1208788; 1208794; 1208795; 1209513-1209517; 1209527-1209531; 1209539; 1209540; 1209546; 1209547; 1210550; 1210551; 1210556-1210558; 1210560; 1210570-1210574; 1210581-1210583; 1210587; 1210590; 1210591; 1211599; 1211602-1211605; 1211608; 1211611; 1211614; 1211616; 1211618; 1211621; 1211623; 1211626; 1211627; 1211635; 1211636; 1211646-1212649; 1212654; 1212655; 1212659; 1212676-1212680; 1212687-1212689; 1212691-1212695; 1301702-1301705; 1301712; 1301713; 1301723-1301725; 1301733-1301735; 1302761-1302763; 1302768; 1302776; 1302777; 1303504-1303508; 1303517; 1303518; 1303522-1303524; 1303785-1303787; 1303794; 1303796; 1303797 & 1304534-1304537. Model #, 110717, Lot #, 1208503; 1208785; 1208786; 1208789; 1208790; 1208796; 1208797; 1208798; 1209504 - 1209510; 1209518-1209521; 209526; 1209533; 1209534; 1209537; 1209541; 1210552; 1210553; 1210559; 1210561; 1210569; 1210575-1210579; 1210584-1210586; 1210589; 1210600; 1211598; 1211601; 1211606; 1211609; 1211610; 1211617; 1211619; 1211620; 1211628-1211630; 1211639-1211642; 1212650-1212652; 1212671-1212673; 1212681-1212685; 1212696; 1212698; 1212699; 1301709-1301711; 1301716-1301719; 1301722; 1301726-1301729; 1301732; 1302754-1302756; 1302766; 1302767; 1302771-1302773; 1302778-1302781; 1303500-1303502; 1303509-1303513; 1303519-1303521; 1303782; 1303788-1303792; 1303798; 1303799; 1304526- 1304531 & 1304542-1304544. Model #, 110718, Lot #, 1208502; 1208782-1208784; 1208792; 1208793; 1208799; 1209511; 1209512; 1209522; 1209523; 1209532; 1209535; 1209536; 1209538; 1209542; 1209544; 1209545; 1210554; 1210555; 1210562-1210568; 1210580; 1210592-1210596; 1211607; 1211612; 1211622; 1211631-1211633; 1211643-1211645; 1212653; 1212674; 1212675; 1212686; 1212700; 1212701; 1301706; 1301708; 1301714; 1301715; 1301720; 1301721; 1301730; 1301731; 1302759; 1302760; 1302764; 1302765; 1302774; 1303503; 1303514-1303516; 1303525; 1303783; 1303784; 1303793; 1304532; 1304533 & 1304539-1304541. Model #, 110721, Lot #, 1210555
Distribution pattern: Worldwide Distribution - USA (nationwide) and internationally to AUSTRALIA, BELGIUM, CANADA, COTE D'IVOIRE, DENMARK, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, IRELAND, ITALY, KENYA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, POLAND, QATAR, RUSSIAN FEDERATION, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM, VIET NAM & YEMEN. .
Derived failure modes
-
Unknown
During a complaint investigation HemoCue AB discovered punctured HemoCue Glucose 201 single pack pouches.