Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer

Z-1270-2014

February 18, 2014

Class III

Product summary

Firm
Lin-Zhi International Inc
Event
Event 67607
Status
Terminated
Classification
Class III
Quantity
120 units, 60 of each lot number
Official record key
device-enforcement:Z-1270-2014

Official wording

Reason: Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).

Code information: Product catalog number: A68827; Lots 1308006, 1308131, Expiry: March 28, 2015.

Distribution pattern: US Distribution including CA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).